This page is part of the FHIR Specification (v1.4.0:
STU
3 Ballot 3). The current version which supercedes this version is
5.0.0
.
For
a
full
list
of
available
versions,
see
the
Directory
of
published
versions
. For a full list of available versions, see the
Directory of published versions
.
Page
versions:
. Page versions:
R5
R4B
R4
R3
R2
Raw JSON
(
canonical
form
canonical form
)
Blood
Pressure
Measurement
)
Blood Pressure Measurement
{
"resourceType": "Observation",
"id": "blood-pressure",
"meta": {
"lastUpdated": "2014-01-30T22:35:23+11:00"
},
"text": {
"fhir_comments": [
" \tEH Narrative created by build\n\t<text>\n\t\t<status value=\"generated\"/>\n\t\t<div xmlns=\"http://www.w3.org/1999/xhtml\">Sept 17, 2012: Systolic Blood pressure 107/60 mmHg (low)</div>\n\t</text>\n "
],
"status": "generated",
"div": "<div><p><b>Generated Narrative with Details</b></p><p><b>id</b>: blood-pressure</p><p><b>meta</b>: </p><p><b>identifier</b>: urn:uuid:187e0c12-8dd2-67e2-99b2-bf273c878281</p><p><b>status</b>: final</p><p><b>code</b>: Blood pressure systolic & diastolic <span>(Details : {LOINC code '55284-4' = 'Blood pressure systolic and diastolic', given as 'Blood pressure systolic & diastolic'})</span></p><p><b>subject</b>: <a>Patient/example</a></p><p><b>effective</b>: 17/09/2012</p><p><b>performer</b>: <a>Practitioner/example</a></p><p><b>interpretation</b>: low <span>(Details : {http://hl7.org/fhir/v2/0078 code 'L' = 'Low', given as 'Below low normal'})</span></p><p><b>bodySite</b>: Right arm <span>(Details : {SNOMED CT code '368209003' = '368209003', given as 'Right arm'})</span></p><blockquote><p><b>component</b></p><p><b>code</b>: Systolic blood pressure <span>(Details : {LOINC code '8480-6' = 'Systolic blood pressure', given as 'Systolic blood pressure'}; {SNOMED CT code '271649006' = '271649006', given as 'Systolic blood pressure'}; {http://acme.org/devices/clinical-codes code 'bp-s' = '??', given as 'Systolic Blood pressure'})</span></p><p><b>value</b>: 107 mm[Hg]</p></blockquote><blockquote><p><b>component</b></p><p><b>code</b>: Diastolic blood pressure <span>(Details : {LOINC code '8462-4' = 'Diastolic blood pressure', given as 'Diastolic blood pressure'})</span></p><p><b>value</b>: 60 mm[Hg]</p></blockquote></div>"
"div": "<div><p><b>Generated Narrative with Details</b></p><p><b>id</b>: blood-pressure</p><p><b>meta</b>: </p><p><b>identifier</b>: urn:uuid:187e0c12-8dd2-67e2-99b2-bf273c878281</p><p><b>status</b>: final</p><p><b>category</b>: Vital Signs <span>(Details : {http://hl7.org/fhir/observation-category code 'vital-signs' = 'Vital Signs', given as 'Vital Signs'})</span></p><p><b>code</b>: Blood pressure systolic & diastolic <span>(Details : {LOINC code '55284-4' = 'Blood pressure systolic and diastolic', given as 'Blood pressure systolic & diastolic'})</span></p><p><b>subject</b>: <a>Patient/example</a></p><p><b>effective</b>: 17/09/2012</p><p><b>performer</b>: <a>Practitioner/example</a></p><p><b>interpretation</b>: Below low normal <span>(Details : {http://hl7.org/fhir/v2/0078 code 'L' = 'Low', given as 'low'})</span></p><p><b>bodySite</b>: Right arm <span>(Details : {SNOMED CT code '368209003' = 'Right upper arm structure (body structure)', given as 'Right arm'})</span></p><blockquote><p><b>component</b></p><p><b>code</b>: Systolic blood pressure <span>(Details : {LOINC code '8480-6' = 'Systolic blood pressure', given as 'Systolic blood pressure'}; {SNOMED CT code '271649006' = 'Systolic blood pressure (observable entity)', given as 'Systolic blood pressure'}; {http://acme.org/devices/clinical-codes code 'bp-s' = '??', given as 'Systolic Blood pressure'})</span></p><p><b>value</b>: 107 mmHg<span> (Details: http://unitsofmeasure.org code mm[Hg] = 'mmHg')</span></p></blockquote><blockquote><p><b>component</b></p><p><b>code</b>: Diastolic blood pressure <span>(Details : {LOINC code '8462-4' = 'Diastolic blood pressure', given as 'Diastolic blood pressure'})</span></p><p><b>value</b>: 60 mmHg<span> (Details: http://unitsofmeasure.org code mm[Hg] = 'mmHg')</span></p></blockquote></div>"
},
"identifier": [
{
"system": "urn:ietf:rfc:3986",
"value": "urn:uuid:187e0c12-8dd2-67e2-99b2-bf273c878281"
}
],
"status": "final",
"category": {
"coding": [
{
"system": "http://hl7.org/fhir/observation-category",
"code": "vital-signs",
"display": "Vital Signs"
}
]
},
"code": {
"coding": [
{
"system": "http://loinc.org",
"code": "55284-4",
"display": "Blood pressure systolic & diastolic"
}
]
},
"subject": {
"reference": "Patient/example"
},
"effectiveDateTime": "2012-09-17",
"performer": [
{
"reference": "Practitioner/example"
}
],
"interpretation": {
"fhir_comments": [
" an interpretation offered to the combination observation\n generally, it would only be appropriate to offer an interpretation\n of an observation that has no value if it has \"COMP\" (component)\n observations "
],
"coding": [
{
"system": "http://hl7.org/fhir/v2/0078",
"code": "L",
"display": "Below low normal"
"display": "low"
}
],
"text": "low"
"text": "Below low normal"
},
"bodySite": {
"coding": [
{
"system": "http://snomed.info/sct",
"code": "368209003",
"display": "Right arm"
}
]
},
"component": [
{
"fhir_comments": [
" \tShould have a refrange as well "
],
"code": {
"fhir_comments": [
" \n Observations are often coded in multiple code systems.\n - LOINC provides a very specific code (though not more specific in this particular case)\n - snomed provides a clinically relevant code that is usually less granular than LOINC\n - the source system provides its own code, which may be less or more granular than LOINC\n ",
" that shows the concept. The next two codes only have a LOINC code "
" \n Observations are often coded in multiple code systems.\n - LOINC provides a very specific code (though not more specific in this particular case)\n - snomed provides a clinically relevant code that is usually less granular than LOINC\n - the source system provides its own code, which may be less or more granular than LOINC\n\nthis is shown here to demonstrate the concept of translations within the codeableConcept datatype. The diastolic code below only has a LOINC code \n "
],
"coding": [
{
"fhir_comments": [
" LOINC -code "
],
"system": "http://loinc.org",
"code": "8480-6",
"display": "Systolic blood pressure"
},
{
"fhir_comments": [
" SNOMED CT Codes "
],
"system": "http://snomed.info/sct",
"code": "271649006",
"display": "Systolic blood pressure"
},
{
"fhir_comments": [
" Also, a local code specific to the source system "
],
"system": "http://acme.org/devices/clinical-codes",
"code": "bp-s",
"display": "Systolic Blood pressure"
}
]
},
"valueQuantity": {
"fhir_comments": [
" no standard units used in this example "
],
"value": 107,
"unit": "mm[Hg]"
"unit": "mmHg",
"system": "http://unitsofmeasure.org",
"code": "mm[Hg]"
}
},
{
"fhir_comments": [
" \tShould have a refrange as well "
],
"code": {
"fhir_comments": [
" this codes only has a LOINC code "
],
"coding": [
{
"system": "http://loinc.org",
"code": "8462-4",
"display": "Diastolic blood pressure"
}
]
},
"valueQuantity": {
"fhir_comments": [
" no formal units in this example "
],
"value": 60,
"unit": "mm[Hg]"
"unit": "mmHg",
"system": "http://unitsofmeasure.org",
"code": "mm[Hg]"
}
}
]
}
Usage
note:
every
effort
has
been
made
to
ensure
that
the
examples
are
correct
and
useful,
but
they
are
not
a
normative
part
of
the
specification.
©
HL7.org
2011+.
FHIR
DSTU2
(v1.0.2-7202)
generated
on
Sat,
Oct
24,
2015
07:44+1100.
Links:
Usage note: every effort has been made to ensure that the examples are correct and useful, but they are not a normative part of the specification.